uofa:requiredLevel 2
uofa:achievedLevel 2
uofa:factorStatus assessed
uofa:acceptanceCriteria Goal (b): N=11 construct samples; nominal geometric range only (no intentional variation in screw diameter, rod diameter, polyaxial angle). [Nagaraja §2.4.1.2, Table 1]
uofa:rationale Eleven non-cannulated constructs assembled to nominal dimensions and tested per ASTM F1717. Sample count exceeds typical N≥6 recommended for spinal device testing.
uofa:factorStandard ASME-VV40-2018