CredibilityFactor
Test conditions
https://uofa.net/nagaraja/cou1/factor/test-conditions
Described in
packs/vv40/examples/nagaraja/cou1/uofa-nagaraja-cou1.jsonld
Fields
@type |
CredibilityFactor |
uofa:factorType |
Test conditions |
uofa:requiredLevel |
3 |
uofa:achievedLevel |
1 |
uofa:factorStatus |
assessed |
uofa:acceptanceCriteria |
Goal (c) intended; Goal (a) achieved: single test condition (ASTM F1717 static compression-bending at 25 mm/min, room temperature, dry). Range, repeats, and uncertainty quantification of test conditions limited. [Nagaraja Table 1, §2.4.1.2] |
uofa:rationale |
BELOW TARGET. Authors acknowledge in §4: 'the sample size chosen for experimental testing may not meet a statistically relevant sample size' — encoded as a factor-level shortfall here. The N=11 sample size partly compensates per ASTM spinal device convention. The shortfall is documented in the decision rationale rather than blocking acceptance. |
uofa:factorStandard |
ASME-VV40-2018 |