Fields

@type CredibilityFactor
uofa:factorType Relevance of the validation activities to the COU
uofa:requiredLevel 3
uofa:achievedLevel 3
uofa:factorStatus assessed
uofa:acceptanceCriteria Goal (c): the validation point IS the COU. COU1 evaluates the non-cannulated construct, which is also the experimental construct. Perfect overlap between validation and COU by construction. [Nagaraja §2.2]
uofa:rationale COU1 is the validation activity itself. The bench test specimens are exactly the constructs the FEA predicts. Highest credibility level achievable by definition.
uofa:factorStandard ASME-VV40-2018

Referenced by

Which records point at this identifier, and through which property. This relationship is not written down anywhere in the packages themselves; it is derived by indexing the whole corpus.