CredibilityFactor
Relevance of the validation activities to the cou
https://uofa.net/nagaraja/cou1/factor/relevance-of-the-validation-activities-to-the-cou
Described in
packs/vv40/examples/nagaraja/cou1/uofa-nagaraja-cou1.jsonld
Fields
@type |
CredibilityFactor |
uofa:factorType |
Relevance of the validation activities to the COU |
uofa:requiredLevel |
3 |
uofa:achievedLevel |
3 |
uofa:factorStatus |
assessed |
uofa:acceptanceCriteria |
Goal (c): the validation point IS the COU. COU1 evaluates the non-cannulated construct, which is also the experimental construct. Perfect overlap between validation and COU by construction. [Nagaraja §2.2] |
uofa:rationale |
COU1 is the validation activity itself. The bench test specimens are exactly the constructs the FEA predicts. Highest credibility level achievable by definition. |
uofa:factorStandard |
ASME-VV40-2018 |
Referenced by
Which records point at this identifier, and through which property. This relationship is not written down anywhere in the packages themselves; it is derived by indexing the whole corpus.
| Record | Property | Source |
|---|---|---|
https://uofa.net/nagaraja/cou1 |
uofa:hasCredibilityFactor |
uofa-nagaraja-cou1.jsonld |