CredibilityFactor
Output comparison
https://uofa.net/nagaraja/cou1/factor/output-comparison
Described in
packs/vv40/examples/nagaraja/cou1/uofa-nagaraja-cou1.jsonld
Fields
@type |
CredibilityFactor |
uofa:factorType |
Output comparison |
uofa:requiredLevel |
3 |
uofa:achievedLevel |
3 |
uofa:factorStatus |
assessed |
uofa:acceptanceCriteria |
Goal (c): three QoIs (construct stiffness, yield load, force at 40 mm displacement) compared between FEA and experiment using ASME V&V 20 Level I (E/D ratio) and Level II (E/uval) metrics. [Nagaraja §2.4.1.4, §3] |
uofa:rationale |
Level I E/D values: stiffness 1.5%, yield load 1.1%, force at 40 mm 1.3% — all within the pre-established 10% acceptance criterion. Level II E/uval values within ±1.96 (95% confidence) for all three QoIs. |
uofa:factorStandard |
ASME-VV40-2018 |
Referenced by
Which records point at this identifier, and through which property. This relationship is not written down anywhere in the packages themselves; it is derived by indexing the whole corpus.
| Record | Property | Source |
|---|---|---|
https://uofa.net/nagaraja/cou1 |
uofa:hasCredibilityFactor |
uofa-nagaraja-cou1.jsonld |