@type |
CredibilityFactor |
uofa:factorType |
Equivalency of input parameters |
uofa:requiredLevel |
3 |
uofa:achievedLevel |
3 |
uofa:factorStatus |
assessed |
uofa:acceptanceCriteria |
Goal (c): types and ranges of all inputs (material properties, geometry, boundary conditions, contact) equivalent between FEA model and experimental setup. [Nagaraja Table 1] |
uofa:rationale |
FEA inputs match experimental measurements: ASTM E8 stress-strain curve for Ti-6Al-4V, measured construct geometry from CAD, F1717-prescribed boundary conditions and loading rate. |
uofa:factorStandard |
ASME-VV40-2018 |