schema:name COU1: Non-cannulated construct compression-bending prediction
schema:description FEA predicts construct stiffness, yield force, and force at 40 mm displacement for the non-cannulated 7.5 mm pedicle screw design under ASTM F1717 static compression-bending. Validated by direct comparison to bench testing (N=11 constructs across two laboratories). Model influence MEDIUM (testing confirms results); decision consequence MEDIUM (pedicle screw fracture leads to revision surgery, not death).
uofa:intendedUse Demonstrate that the FEA modeling framework reproduces experimental compression-bending behavior of a non-cannulated pedicle screw construct within pre-established acceptance bounds (<10% QoI difference, E/uval < 1.96).
uofa:deviceClass Class II
uofa:modelInfluence Medium
uofa:decisionConsequence Medium