ContextOfUse
Cou1 noncannulated
https://uofa.net/nagaraja/cou/cou1-noncannulated
Described in
packs/vv40/examples/nagaraja/cou1/uofa-nagaraja-cou1.jsonld
Fields
@type |
ContextOfUse |
schema:name |
COU1: Non-cannulated construct compression-bending prediction |
schema:description |
FEA predicts construct stiffness, yield force, and force at 40 mm displacement for the non-cannulated 7.5 mm pedicle screw design under ASTM F1717 static compression-bending. Validated by direct comparison to bench testing (N=11 constructs across two laboratories). Model influence MEDIUM (testing confirms results); decision consequence MEDIUM (pedicle screw fracture leads to revision surgery, not death). |
uofa:intendedUse |
Demonstrate that the FEA modeling framework reproduces experimental compression-bending behavior of a non-cannulated pedicle screw construct within pre-established acceptance bounds (<10% QoI difference, E/uval < 1.96). |
uofa:deviceClass |
Class II |
uofa:modelInfluence |
Medium |
uofa:decisionConsequence |
Medium |