ContextOfUse
Cou2 vad
https://uofa.net/morrison/cou/cou2-vad
Described in
packs/vv40/examples/morrison/cou2/uofa-morrison-cou2.jsonld
Fields
@type |
ContextOfUse |
schema:name |
COU2: Ventricular assist device use (Class III) |
schema:description |
A CFD model is used to predict absolute hemolysis levels across operating conditions for a centrifugal blood pump used as a ventricular assist device (VAD). Model influence is HIGH (model is primary basis for hemolysis acceptance); decision consequence is HIGH (incorrect prediction could result in chronic patient hemolysis requiring clinical intervention). Model Risk Level 5 — the highest in the Morrison assessment. |
uofa:intendedUse |
Use CFD to predict absolute hemolysis index values for a Class III VAD device across 5 flow-speed combinations. Results must demonstrate hemolysis below absolute clinical threshold (not relative to predicate). Requires Monte Carlo uncertainty quantification on hemolysis predictions. |
uofa:deviceClass |
Class III |
uofa:modelInfluence |
High |
uofa:decisionConsequence |
High |