ContextOfUse
Cou1 cpb
https://uofa.net/morrison/cou/cou1-cpb
Described in
packs/vv40/examples/morrison/cou1/uofa-morrison-cou1.jsonld
Fields
@type |
ContextOfUse |
schema:name |
COU1: Cardiopulmonary bypass use (Class II) |
schema:description |
A CFD model is used to evaluate the sensitivity of pump-induced hemolysis to variations in component dimensions for a centrifugal blood pump used during cardiopulmonary bypass surgery. Model influence is MEDIUM (testing confirms results); decision consequence is HIGH (incorrect tolerance decision could impact clinical hemolysis levels). Model Risk Level 2. |
uofa:intendedUse |
Use CFD to identify worst-case operating conditions for hemolysis across 5 flow-speed combinations for a Class II CPB device. Results compared against predicate device hemolysis using relative hemolysis index (RH < 1 acceptance criterion). |
uofa:deviceClass |
Class II |
uofa:modelInfluence |
Medium |
uofa:decisionConsequence |
High |
Referenced by
Which records point at this identifier, and through which property. This relationship is not written down anywhere in the packages themselves; it is derived by indexing the whole corpus.
| Record | Property | Source |
|---|---|---|
https://uofa.net/morrison/cou1 |
uofa:hasContextOfUse |
uofa-morrison-cou1.jsonld |