Unit of Assurance

UncertaintyQuantification

Monte carlo hemolysis cou2

https://uofa.net/morrison-v09/uq/monte-carlo-hemolysis-cou2

Described in packs/vv40/examples/morrison-v09/cou2/uofa-morrison-v09-cou2.jsonld

Fields

@type UncertaintyQuantification
schema:name Monte Carlo hemolysis UQ
schema:description 1000-sample Monte Carlo propagation of input uncertainties (blood viscosity, inlet boundary conditions, hemolysis model constants) through CFD pipeline. 95% confidence intervals on predicted MIH at each of 5 operating conditions. This is the key structural difference from COU1, which has no UQ on hemolysis predictions.

Referenced by

Which records point at this identifier, and through which property. This relationship is not written down anywhere in the packages themselves; it is derived by indexing the whole corpus.

RecordPropertySource
https://uofa.net/morrison-v09/claim/cou2-hemolysis-adequacy prov:wasDerivedFrom uofa-morrison-v09-cou2.jsonld
https://uofa.net/morrison-v09/validation/hemolysis-comparison-cou2 uofa:hasUncertaintyQuantification uofa-morrison-v09-cou2.jsonld