Unit of Assurance

DecisionRecord

Cou2 overall

https://uofa.net/morrison-v09/decision/cou2-overall

Described in packs/vv40/examples/morrison-v09/cou2/uofa-morrison-v09-cou2.jsonld

Fields

@type DecisionRecord
uofa:role Credibility assessment team
uofa:outcome Not accepted
uofa:rationale Model not deemed sufficiently credible for COU2. All seven assessed credibility factors have achieved levels below the required levels for Model Risk Level 5. Key gaps: software quality assurance (achieved 1 vs required 3), numerical code verification (achieved 3 vs required 4), discretization error (achieved 3 vs required 4), test conditions (achieved 2 vs required 4), output comparison (achieved 2 vs required 4), relevance of quantities of interest (achieved 2 vs required 3), and relevance of validation activities to the COU (achieved 2 vs required 3). Additionally, 6 of 13 credibility factors were not assessed, representing epistemic gaps at MRL 5. The evidence shows the model does not meet the credibility bar for a Class III VAD application. Additional verification and validation activities required before the model can support Class III VAD hemolysis claims.
uofa:decidedAt 2019-02-01T00:00:00Z
uofa:decisionProvenance extracted
uofa:decisionAnchor
{"type":"SourceAnchor","anchorLocator":"archive://morrison/source/decision_rationale_cou2.pdf","anchorSha256":"sha256:e64fc31cc0ccbb4e88c6bbed65d4ae2d24e7f8f25c428dda62607bed5fd6f4a9"}

References

uofa:actor https://uofa.net/org/FDA-CDRH

Referenced by

Which records point at this identifier, and through which property. This relationship is not written down anywhere in the packages themselves; it is derived by indexing the whole corpus.

RecordPropertySource