Fields

@type ContextOfUse
schema:name COU2: Ventricular assist device use (Class III)
schema:description A CFD model is used to predict absolute hemolysis levels across operating conditions for a centrifugal blood pump used as a ventricular assist device (VAD). Model influence is HIGH (model is primary basis for hemolysis acceptance); decision consequence is HIGH (incorrect prediction could result in chronic patient hemolysis requiring clinical intervention). Model Risk Level 5 — the highest in the Morrison assessment.
uofa:intendedUse Use CFD to predict absolute hemolysis index values for a Class III VAD device across 5 flow-speed combinations. Results must demonstrate hemolysis below absolute clinical threshold (not relative to predicate). Requires Monte Carlo uncertainty quantification on hemolysis predictions.
uofa:deviceClass Class III
uofa:modelInfluence High
uofa:decisionConsequence High