schema:name COU1: Cardiopulmonary bypass use (Class II)
schema:description A CFD model is used to evaluate the sensitivity of pump-induced hemolysis to variations in component dimensions for a centrifugal blood pump used during cardiopulmonary bypass surgery. Model influence is MEDIUM (testing confirms results); decision consequence is HIGH (incorrect tolerance decision could impact clinical hemolysis levels). Model Risk Level 2.
uofa:intendedUse Use CFD to identify worst-case operating conditions for hemolysis across 5 flow-speed combinations for a Class II CPB device. Results compared against predicate device hemolysis using relative hemolysis index (RH < 1 acceptance criterion).
uofa:deviceClass Class II
uofa:modelInfluence Medium
uofa:decisionConsequence High