  "id": "https://uofa.net/nagaraja/cou1/factor/test-conditions",
  "type": "CredibilityFactor",
  "factorType": "Test conditions",
  "requiredLevel": 3,
  "achievedLevel": 1,
  "factorStatus": "assessed",
  "acceptanceCriteria": "Goal (c) intended; Goal (a) achieved: single test condition (ASTM F1717 static compression-bending at 25 mm/min, room temperature, dry). Range, repeats, and uncertainty quantification of test conditions limited. [Nagaraja Table 1, §2.4.1.2]",
  "rationale": "BELOW TARGET. Authors acknowledge in §4: 'the sample size chosen for experimental testing may not meet a statistically relevant sample size' — encoded as a factor-level shortfall here. The N=11 sample size partly compensates per ASTM spinal device convention. The shortfall is documented in the decision rationale rather than blocking acceptance.",
  "factorStandard": "ASME-VV40-2018"
}