  "id": "https://uofa.net/morrison/validation/piv-velocity-comparison",
  "type": "ValidationResult",
  "name": "PIV velocity field comparison",
  "description": "CFD-predicted velocity fields compared against particle image velocimetry (PIV) measurements at 5 flow-speed operating conditions. Qualitative agreement in flow structures; quantitative comparison of velocity magnitude profiles at selected planes.",
  "outcome": "pass",
  "wasDerivedFrom": "https://uofa.net/morrison/comparator/piv-experimental",
  "wasGeneratedBy": {
    "id": "https://uofa.net/morrison/activity/piv-comparison",
    "type": "VerificationActivity",
    "name": "PIV velocity validation",
    "description": "Comparison of CFD velocity predictions against experimental PIV measurements at matched operating conditions."
  },
  "comparedAgainst": {
    "id": "https://uofa.net/morrison/comparator/piv-experimental",
    "type": "Comparator",
    "name": "PIV experimental velocity data",
    "description": "Particle image velocimetry measurements of velocity field in the FDA generic blood pump at 5 operating conditions. Data collected by FDA OSEL."
  },
  "hasUncertaintyQuantification": {
    "id": "https://uofa.net/morrison/uq/piv-velocity-uncertainty",
    "type": "UncertaintyQuantification",
    "name": "PIV comparison uncertainty",
    "description": "PIV measurement uncertainty propagated through comparison metric. CFD prediction uncertainty from Monte Carlo input perturbation included in comparison bounds."
  }
}