  "id": "https://uofa.net/morrison-v09/uq/monte-carlo-hemolysis-cou2",
  "type": "UncertaintyQuantification",
  "name": "Monte Carlo hemolysis UQ",
  "description": "1000-sample Monte Carlo propagation of input uncertainties (blood viscosity, inlet boundary conditions, hemolysis model constants) through CFD pipeline. 95% confidence intervals on predicted MIH at each of 5 operating conditions. This is the key structural difference from COU1, which has no UQ on hemolysis predictions."
}