  "id": "https://uofa.net/morrison-v09/decision/cou1-overall",
  "type": "DecisionRecord",
  "actor": "https://uofa.net/org/FDA-CDRH",
  "role": "Credibility assessment team",
  "outcome": "Accepted",
  "rationale": "Model deemed sufficiently credible for COU1. All verification goals achieved at or above target gradation levels. Validation comparison at all 5 operating conditions shows hemolysis index below CPB predicate threshold (RH < 1). No formal CFD uncertainty quantification performed for COU1 (addressed by W-AL-01 weakener).",
  "decidedAt": "2019-02-01T00:00:00Z",
  "decisionProvenance": "extracted",
  "decisionAnchor": {
    "type": "SourceAnchor",
    "anchorLocator": "archive://morrison/source/decision_rationale_cou1.pdf",
    "anchorSha256": "sha256:01190eb07629faf0b83aa29c3dd2829d599137f70f1f9446c87c1970bb7eb211"
  }
}